Product Library

How to calibrate Capnoxi Capnography Monitor

CapnOxi  meditech Capnograph

Calibrating the CO2 Measurement
It is recommended to calibrate CO2 measurement when you start CO2 monitoring. But you must carry out the CO2 calibration when the CO2 need zero calibration displays in the alarm area.
 
Procedure:
1. Press menu key on the keyboard to enter the Main Menu dialog.
2. Select User Settings button to enter the User Settings dialog.
3. Select Calibrate Zero button to calibrate CO2 measurement.
4. About 25 seconds later, if the calibration failed, the failure message will be shown.
 

Promoted Products

Meditech Brands

Meditech Equipment Co.,Ltd is part of Meditech Group. Product(s) described may not be licensed or available for sale in all countries. Sonotech, Sonovet, iSonic, FOs2pro, Dolphi, Defi, HeartRec,miniScan,Cardios,SpirOx,iBreath, Meditech and all corresponding design marks are trademarks of Meditech. The symbol indicates the trademark is registered. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. Please see the Instructions for Use for a complete listing of the indications, contraindications, warnings and precautions.

Legal notice Terms and conditions Cookie policy Privacy Policy Professional organisations Careers

To Top