Product Library

SPIROX-P is intended for monitoring PEF (Peak Expired Flow R





Intended Use
SPIROX-P is intended for monitoring PEF (Peak Expired Flow Rate) and FEV1 (Forced Expiratory Volume in One Second) for patient use at home, clinics and hospital. The device is designed for pediatric to adult patients. The simple device interface provides ease Of use for pediatric patients. When the Meter is used to monitor lung conditions such as asthma or COPD, the user should be under the care of a licensed health care professional. A licensed health care professional's advice is required to understand the meaning and importance of The measured reported by the Meter and how to decide on an appropriate treatment plan. This treatment plan will tell the patient what action to take when there are changes in their peak flow measurements. Intended user of SPIROX-P is patients.
 



1 TECHNICAL SPECIFICATION
Sensor type Turbine
Measurement parameters PEF / FEV1
Memory size 250
BTPS PEF and FEV1 results displayed at BTPS conditions
Accuracy
Volume range 0.01~9.99L Air flow range 50 L/min ~900L/min
Volume accuracy ±0.05L or ±3% Air flow accuracy ±10% or ±20L/min
Display screen
Size 1.8 inches
Type LCD
Wireless
Wireless technology Bluetooth Low Energy (4.0)
Output Watts 0.0003
Frequency Range (MHz) 2402.0 – 2480.0
Modulation GFSK
Environment specification
Power source  AAA1.5×2 batteries
Operation Temperature
Humidity
Atmosphere
Altitude limit
 
10 o C~40 o C
25%~85%
860hPa~1060hPa
No limit
Storage
Temperature
Humidity
Atmosphere
 
-20 o C~55 o C
15%~93%
500hPa~1060hPa
Physical specification
Height 12.6 centimeter
Wide 5.4 centimeter
Thickness 2.8 centimeter
Weight 120g(contain batteries and mouthpiece)
Durable Period
Main Unit 3 years
Turbine 2 year
 


Promoted Products

Meditech Brands

Meditech Equipment Co.,Ltd is part of Meditech Group. Product(s) described may not be licensed or available for sale in all countries. Sonotech, Sonovet, iSonic, FOs2pro, Dolphi, Defi, HeartRec,miniScan,Cardios,SpirOx,iBreath, Meditech and all corresponding design marks are trademarks of Meditech. The symbol indicates the trademark is registered. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. Please see the Instructions for Use for a complete listing of the indications, contraindications, warnings and precautions.